Laboratory NES713 Smoke Toxicity Index Test Chamber with Burning 100g Specimen
  • Product Details:
  • Place of Origin: China
  • Brand Name: YUYANG
  • Certification: NES 713 JB/T 10707
  • Model Number: NES 713 JB/T 10707
  • Payment & Shipping Terms:
  • Minimum Order Quantity: 1 set
  • Price: Negotiation
  • Packaging Details: Wood case
  • Delivery Time: 20 work days
  • Payment Terms: T/T L/C Western Union
  • Supply Ability: 1 set per month
  • Share to£º
Product Details

Laboratory NES713 Smoke Toxicity Index Test Chamber with Burning 100g Specimen

 

 

Description:
 
This tester measures the Toxicity Index by ppm in Naval Engineering Standard chamber when complete combustion occurs by burner.
NES 713 TOXIC Chamber meets the standard of NES 713 and is used to measure combustion characteristics about materials.
NES 713 Toxic Chamber measures Toxicity Index after perfect combustion by burning 100g of specimen.
 
 
Standard:
 
NES 713: Determination Of the Toxicity Index Of the Products of Combustion From Small Specimen of Materials
JB/T 10707: Low-Smoke Halogen-free Thermoplastic Compounds for Flame Retardant Cables
 
 
Feature:
 
1. Observation window made in reinforced plastic allows easy observation of combustion conditions during the test
2. Treating the inside of Chamber with Teflon Coating minimizes chemical reaction with gaseous materials generated in combustion.
3. Door locking device designed for easy opening/closing.
4. Specimen Support for Burner to be located in the center of the combustion mass and for test specimen to be easily loaded
5. Designed for testing in Auto/Manual Mode
6. Auto Ignition for controlling ignition and extinguishment of Gas Burner by Time Device
7. Auto Vent Mode allowing auto release of combustion gases generated in combustion by Time Device
8. Auto Stirrer Mode for complying with conditions of combustion gas stirrer by Time Device specified in the standard Time Device
 
 
Application:
 
The NES 713 test explores the toxicity of the products of combustion in terms of small molecular species arising when a sample of a material is completely burnt in excess air under specified conditions. The test does not necessarily determine the total toxicity of all the constituents of the products of combustion.
The test is useful for the quality control of materials and for research and development. It may be used to compare the particular combustion characteristic of a series of both natural and synthetic materials. The test may be used to specify a quality of a raw material or product. Combustion characteristics tests alone are not suitable for assessing the total fire hazard of products under actual fire conditions.
The toxicity index is defined as the numerical summation of the toxicity factor of selected gases produced by complete combustion of the material in air under the conditions specified. The toxicity factors are derived from the calculated quantity of each gas that would be produced when 100g of the material is burnt in air in a volume
 
 
Specification:
 
Model YY435
Size 1,185(W)×700(D)×2,100(H) mm
Power AC 220V, 50/60Hz, 10A
Weight About 150kg
Instructions Supplied
Exhaust 50L/s
Tool Compressed air, vacuum cleaners, methane, propane gas (optional)
 
 
Test Result:
 
If 100g of material is burned and the material is diluted in the air of chamber, concentration of each generated gas is given by the following equation.
C × 100 ×V
C8 = --------------- (ppm)
m
C = concentration of gas in test chamber (ppm)
m = fire test mass (g)
V = volume of test chamber (m3)
 
Toxicity Index
 
C81 C82 C83 C8n
Toxicity Index = Σ ------ + ------ + ------ + … + ------
Cf1 Cf2 Cf3 Cfn
1, 2, 3, ... indicates each found Gas.
Cf = concentration of the gas considered fatal to man for a 30 minute exposure time(ppm)
 
Toxicity Constitution& Concentration ( ppm )
Carbon dioxide(CO2) 100,000
Carbon monoxide(CO) 4,000
Hydrogen sulphide(H2S) 750
Ammonia(NH3) 750
Formaldehyde(HCHO) 500
Hydrogen chloride(HCl) 500
Acrylonitrile(CH2CHCN) 400
Sulphur dioxide(SO2) 400
Nitrogen oxides(NO+NO2) 250
Phenol(C6H5OH) 250
Hydrogen cyanide(HCN) 150
Hydrogen bromide(HBr) 150
Hydrogen fluoride(HF) 100
Phosgene(COCl2) 25
 
Laboratory NES713 Smoke Toxicity Index Test Chamber with Burning 100g Specimen