UV Accelerated Weathering Test Chamber is applicable for the sunlight resistant test of nonmetallic materials and aging test of artificial light sources. A variety of industrial products can make the reliability test, and SUS 304 Steel UVB Accelerated Weathering Testing Chambers can simulate the product in the sun, rain, humidity and dew conditions, including damage caused by bleaching, color, brightness down, powder, crack, blur, brittle, intensity decreased and oxidation.
Technical Parameter:
Model |
HH1004 |
Test chamber size(L×W×H)MM |
1170×450×500 |
Outside dimension(L×W×H)MM |
1350×540×1480 |
Chamber materials |
304# stainless steel |
Temperature range |
RT+20¡æ~70¡æ |
Temperature uniformity: |
±1¡æ |
Temperature fluctuation: |
±0.5¡æ |
Temperature control |
PID SSR control |
Humidity range |
≥90%RH |
Controller |
TEMI 880 programmable controller, touch screen, LCD display |
Control mode |
Balance temperature humidity control (BTHC) |
Communication function |
Be able to control the machine via computer using TEMI control software through RS-232 port on the machine |
Test cycle setting |
Illumination, condensation and water spray test cycle is programmable |
Distance from specimen to lamp |
50±3mm (adjustable) |
Center distance between the lamps |
70mm |
|
Lamp power & length |
40W/Piece, 1200mm/Piece |
|
Lamps amount |
8 pieces of UVA-340 or UVB-313 imported lamps |
|
Lifetime of lamp |
1600Hours |
|
Irradiance |
1.0W/m2, adjustable directly on TEMI 880 touch screen controller |
|
Wavelength of ultraviolet light |
UVA is 315-400nm and UVB 280-315nm |
|
Effective irradiation area |
900×210mm |
|
Irradiation black panel temperature |
50¡æ~70¡æ |
|
Standard specimen size |
75×290mm/24 pieces |
|
Water depth for water channel |
25mm, automatically control |
|
Water consumption |
8L/day, tap water or distilled water |
|
Testing time |
0¡«999H, adjustable |
|
Power |
1 phase, 220V/50Hz /±10% 5KW |
|
Protection |
Overload short circuit protection, over temperature protection, water lacking protection |
|
Chamber Structure:
The inner bladder of the chamber is made of imported high-class stainless steel plate while the outer bladder is made of SUS 304 steel plate.
The heating mode adopts inner bladder trough type heating with its rapid temperature rise and even temperature distribution.
Control System:
Adopts black aluminum plate to connect temperature sensor and adopts black board temperature meter to control heating so as to ensure more stable temperature.
The radiometer probe is fixed so as to avoid the frequent installation and disassembly.
The radiation quantity adopts the special UV irradiatometer with high-precision display and measurement.
The radiation intensity is not more than 50W/m2.
The illumination and condensation can be controlled independently or alternating and circularly.
Test Specimen Mounting:
The standard test specimen holder accommodates 2 panels 48 pcs 75×150mm and 15 pcs 100*300mm,. The test panels are actual the side wall of the chamber. To properly seal the chamber, it is important to have every panel holder filled with specimens or blank panels and to have panel holders in place.
Design Standards:
(1) ASTM D4587: Standard Practice for Fluorescent UV-Condensation Exposures of Paint and Related Coatings
(2) ASTM D4799: Standard Practice for Accelerated Weathering Test Conditions and Procedures for Bituminous Materials (Fluorescent UV, Water Spray, and Condensation Method)
(3) ISO 4892-1: Plastics - Methods of Exposure to Laboratory Light Sources - Part 1: General guidance
(4) ISO 4892-3: Plastics - Methods of Exposure to Laboratory Light Sources - Part 3: Fluorescent UV Lamps
(5) ASTM D4329: Standard Practice for Fluorescent UV Exposure of Plastics
(6) AATCC TM186: Weather Resistance: UV Light and Moisture Exposure
(7) ASTM D5208: Standard Practice for Fluorescent Ultraviolet (UV) Exposure of Photodegradable Plastics
(8) IEC 61345: UV Test for Photovaltalic (PV) Modules